IP54 (NEMA 12) with standard connector, (IP66, NEMA 4X) with optional moulded connector
Range
Calibrated from -100 to +20°C dew point
Operating Temperature
-40 to +60°C (compensated over -20 to +40°C)
Accuracy
±1°C from -59 to +20°C dew point; ±2°C from -100 to -60°C dew point
Resolution
0.1°C from -80 to +20°C dew point; 1.0°C from -100 to -80°C dew point
INSTRUMENT
Measurement Units
°C, °F dew point; ppmV with user selectable range and resolution; Lb/mmscf range dependent on pressure (natural gas); gm-³ (natural gas)
Secondary Measurement Variable
Pressure via external transducer 0-30 MPa Completely user configurable for any industry standard transducer Automatic compensation for ppmV, Lb/mmscf. and gm-³ ranges
Power Supply
Universal 85 to 265 VAC; 95 to 370 VDC
Ingress Protection
IP54 (NEMA 4); IP66 (NEMA 4X) for front panel only with optional cover
Operating Temperature
-20 to +50°C
Relays
2 off 10 A/240 V c/o fully user configurable: can accommodate full range for selected engineering unit plus sensor fault alarm functions
Outputs
Analogue: 4-20 mA fully user configurable and scaleable for range Digital: RS232: provides 2-way communication-full data access to user plus total availability of all set-up and configuration
OPTIONS:
Power Supply
18 to 36 VAC; 9 to 60 VDC
Relays
1 or 2 further 5A relays (max 4 in total) n/o type fully user configurable. Can accommodate full range + full unit choice + sensor fault alarm functions
Outputs
Analogue: 0-10 V fully user configurable and scaleable for range Digital: range RS485: provides 2-way communication, full data access to user + total availability of all set-up and configuration