The S8000 Integrale continues Michell's tradition of excellence in the design of high performance humidity products. Building from the solid foundations of performance and reliability in its predecess
详细介绍
Product Features Fundamental, accurate and drift free measurement 1 mK display resolution -60 to +40°C dew-point range with 0.1°C accuracy FAST-guaranteed frost formation below 0°C Data logging to USB or SD card Multi-language display Vertical or horizontal configuration
Technical Specification
Measurement Range
-60°C to +40°C dew point
Measurement Accuracy
±0.1°C dew point
Stability
<0.05°C
Repeatability
<0.05°C
Measurement Units
°C, °F dew point; % RH; °C, °F temperature; gm-³, g/kg-¹, ppmV; ppmW; SF6; ml/min-¹
Response Speed
2°C sec-¹ + settling time
Power Supply
85 to 264 V, 47/63 Hz; 100 VA
Operating Temperature Range
-20 to +50°C
DEW POINT SENSOR
Mirror
Gold plated copper
Temperature Measurement
4-wire 100 Ω 1/10 DIN class B Pt100
Pressure
17 Barg (for higher pressure consult factory)
FLOW SENSOR
Measurement Range
0-1000 ml/min
DISPLAY UNIT
Resolution
User-selectable to 0.001 dependent on parameter
Outputs
Analogue: Three channels; user-selectable 4-20 mA, 0-20 mA or 0-1V Digital: USB, SD Card slot Alarm: Two volt free changeover contacts; 1A @ 30VDC
Cable Pack
Supply and USB cables. Output connectors suitable for indoor use
OPTIONAL REMOTE PRT
Temperature Measurement
4 wire 100 Ω 1/10 DIN class B Pt100
Measurement Accuracy*
±0.1°C
Measurement Units
°C or °F
Cable Units
2m (250m max)
OPTIONAL PRESSURE SENSOR (BUILT-IN)
Measurement Range
0 to 25 Barg
Measurement Accuracy*
0.25% Full Scale, typical
Measurement Units
Bara; Psia; Kpa
OPTIONS:
Microscope
Rack-mount kit (horizontal version)
UKAS Calibration
Transportation case
*Measurement accuracy means maximum deviation between instrument under test and corrected reference. To this must be added the uncertainties associated with the calibration system and the environmental conditions during testing or subsequent use.