供应 4OXV /7OX-V氧气传感器4OX(2)
4OXV Oxygen Sensor CiTiceL ®
7OX-V Oxygen Sensor CiTiceL ®
CiTiceL® O2 4OXV 2 years in air 0-30%
Technical Specifications
Electrochemical
1-25% vol. O 2
30% vol. O 2
0.10 ± 0.02 mA in Air
<15 Seconds
<0.6% vol. O 2
Can be considered linear in
many cases.
See Operating Principles
(OP-02) for further details.
MEASUREMENT
Technology
Measurement Range
Maximum Overload
Output Signal*
Response Time (T90)*
Zero Current (Offset)*
Linearity
100 Ω Recommended Load Resistor
ABS
<16 g
<0.2% vol. O 2 equivalent
MECHANICAL
Casing Material
Weight
Orientation Sensitivity
-20°C to +50°C (up to 3 months
continuous across RH range)
0°C to 20°C
<23.5% vol. O 2
Atmospheric ± 20%
<0.02% signal/mbar
<200% signal change
(at 0°C to 20°C)
5 to 95%RH non-condensing
0 to 99%RH non-condensing
ENVIRONMENTAL
Operating Temperature Range
Recommended Storage Temp
Thermal Transient
(Temp. plunge +22°C to -20°C)
Operating Pressure Range
Pressure Coefficient
Pressure Transient
(60 cm H 2 O step change)
Relative Humidity Range
Continuous
Short Term
<2% signal/month
Typically <5% over operating life
Minimum 24 months in air
6 months in original packaging
24 months from date of despatch
LIFETIME
Long Term Output Drift*
Expected Operating Life
Storage Life
ELECTRICAL
IMPORTANT NOTES
All tolerances ±0.15 mm unless otherwise stated.
Do not remove label. Do not solder to pins.
When installing the sensor into instrumentation,
the sensor vent hole should not be blocked. The
instrument should also be adequately vented.
If the sensor vent hole is blocked or if the instrument
is not adequately vented, sensor performance will be
compromised.