Measurement Range | RS: -85 to +20°C TRS: -100 to +20°C |
Measurement Accuracy | ±0.1°C dew point ±0.1°C temperature |
Measurement Units | °C, °F dew point; °C, °F temperature; % RH, ppmV, ppmW, gm-³, g/kg-¹, ppmW for SF6 |
Response Speed | 0.5°C s-¹ + settling time (dew-point dependent) |
Sensitivity | 0.01°C |
Repeatability | Better than 0.1°C |
Resolution | 0.01 (0.1 for % RH) |
DEW POINT SENSOR |
Mirror | Gold plated copper |
Temperature Measurement | 4 wire 100 Ω 1/10 DIN class B Pt100 |
Sample Flow Rate | 0 to 1 Nl/min (recommended) |
Integrated Flowmeter | 0 to 1 Nl/min |
Sensor Pressure | 1 MPa (10 Barg) max |
Auxiliary Cooling | Internal refrigeration |
REMOTE PRT |
Temperature Measurement | 4 wire 100 Ω 1/10 DIN class B Pt100 |
MONITOR |
Resolution | 0.01°C |
Dual Optics Detection | Wide band red LED with dual photo sensors, all systems insulated |
Outputs | Analogue: 2 channels-10mV/°C dew point, 4-20 mA Digital: RS-232 Logic: data hold ABC logic status, optics alarm |
Auxiliary Input | Pressure transducer: 4-20 mA input for automatic compensation 0-50 psia (optional) |
Operating Temperature | 0 to +40°C |
Dimensions | RS: 482W x 510D x 402H mm-Sub Rack TRS: 560W x 600D x 860H mm-Mini Rack |
Weight | RS: 32.4kg TRS: 85.0kg |
Power Supply | Monitor: 90 to 265 VAC; 50-60 Hz Sensor: 100-115 or 220-240 VAC; 50-60 Hz |